|
Newer Modes of Mechanical Ventilation
Recently, modes of mechanical ventilation that synchronise
not only the timing, but also the level of assist to the patient's own effort,
have been introduced
"Microprocessor
based technology has led to the development of newer modes, aimed at improving
ventilatory support"
- Dr Deepak Govil
Consultant Intensivist
Critical Care
Artemis Health Institute
New Delhi
|
Mechanical ventilation, originally used for resuscitation,
consisted of simple, manually-driven pump devices. With the advent of positive
pressure ventilators, mechanical ventilation has become an integral part of
the care of patients with acute respiratory failure. The crudest form delivers
fixed volumes or pressures with fixed rates and breath durations, while more
advanced modes allow the patient to trigger the ventilator with their own effort,
which can then be cycled off when certain criteria have been met, attempting
to mimic natural breathing.
The primary goals of mechanical ventilation is to 'buy time'
to give patient a chance to recover from some catastrophe by providing adequate
gas exchange and unloading the respiratory muscles without causing any further
damage to respiratory muscles or lung parenchyma.
The early re-institution of spontaneous breathing in mechanically-ventilated
critically ill patients becomes an important therapeutic option to avoid the
various complications of controlled mechanical ventilation and facilitate the
weaning process. In spontaneously breathing patients, it is also important to
achieve patient-ventilator synchrony, that is, the ventilator needs to cycle-on
in unison with the onset of the patient's effort. The level of assist delivered
should correspond to the patient's demand and most importantly, the ventilator
breath should be cycled-off at the end of the patient's inspiratory effort.
Co-ordination between spontaneous breathing and mechanical assistance may not
be achieved always and a poor interaction between patient and machine may cause
either excessive or insufficient ventilatory support or dys-synchrony. Microprocessor-based
technology has led to the development of newer modes aimed at improving ventilatory
support.
ASV
Adaptive
Support Ventilation (ASV) is a pressure controlled closed-loop system which
allows adaptation of assistance during all phases of mechanical ventilation,
from control ventilation to weaning. The main principle of functioning is based
on Otis's formula which is stored in microprocessor of ventilator. Using this
formula, the ventilator can calculate an 'ideal' respiratory pattern (tidal
volume and respiratory rate), which needs the smallest total energy expenditure,
providing specific minute ventilation, a calculated dead space, which depends
on body weight and an expiratory time constant. Minimum minute ventilation is
the only specific setting that must be chosen by the clinician. It is based
on patient's body weight.
When starting ventilation in ASV, the ventilator provides three pressure control
time cycle inspirations and calculates respiratory mechanics. Expiratory time
constant is estimated from the tidal volume curve during each expiration. Then,
using Otis' formula, a target respiratory rate is calculated. Target tidal volume
is computed from minimum minute ventilation and target respiratory rate. Thereafter,
target values are calculated cycle-by-cycle. Depending on patient's spontaneous
respiratory rate, ASV can works as Pressure Control Ventilation (PCV), if there
is no spontaneous breathing, as pressure control SIMV, when patient's respiratory
rate is smaller than target, or as Pressure Support Ventilation (PSV), if patient's
respiratory rate is greater. Pressure level is then adapted to attain the target
tidal volume (within limits imposed by pressure alarms). Cycling off criteria
is flow-based in the case of assisted ventilation or time-based for mandatory
inspiration.
With
ASV, the patient has a high degree of breathing freedom within each breath,
since the breath delivery by ventilator is by controlling the inspiratory pressure
and never by forcing a given inspiratory flow. With ASV, the user-set minimum
minute ventilation is always guaranteed, with a theoretically optimal ventilatory
pattern, while the patient always maintains the freedom of increasing his minute
ventilation above the user-set target.
Several studies have shown that ASV can provide safe and effective ventilation
in patients with normal lungs, restrictive or obstructive diseases. Other benefits
include decrease of patient's respiratory efforts, stability of alveolar ventilation
without operator's intervention and safety during weaning in selected situations.
However, many important unsolved questions remain to be answered for ASV like
how minimum ventilation must be set and how ASV must be adapted faced with changes
in breathing demand and how weaning must be handled?
New Modes
Recently,
modes of mechanical ventilation that synchronise not only the timing, but also
the level of assist to the patient's own effort have been introduced, such as
Proportional Assist Ventilation (PAV) and Neurally Adjusted Ventilatory Assist
(NAVA). PAV and NAVA have been designed with the goal of improving patient-ventilator
interaction by matching the ventilator support with the neural output of the
respiratory centres. With PAV, the support is continuously re-adjusted in proportion
to the predicted inspiratory effort. NAVA is a mode in which the assistance
is delivered in proportion to the electrical activity of the diaphragm, assessed
by means of an esophageal electrode.
PAV: Novel Mode
Proportional assist ventilation (PAV) is a novel mode of partial ventilatory
support in which the ventilator generates an instantaneous inspiratory pressure
in proportion to the instantaneous effort of the patient, that is, the more
the patient pulls, the more pressure the machine generates. Thus, with PAV the
ventilator amplifies the patient's inspiratory effort without any preselected
target volume or pressure. There are two main forces to overcome when breathing
in.
Firstly, the resistance of the airways and secondly the breathing circuit and
the elasticity of the chest wall and lung. In PAV, the support that is applied
at any time point during inspiration is in proportion to the work required to
overcome the individual resistive and elastic forces. To offload the resistive
work, pressure is applied in proportion to the flow rate, and to offload the
elastic work, pressure is applied in proportion to the volume inhaled. To apply
PAV correctly, the resistance and elastance need to be known. The degree of
support applied, is then set as a percentage between zero percent and 90 per
cent. Levels of support above 90 per cent are not applied due to the risk of
amplification or 'runaway.' The aim of PAV is to allow the patient to attain
whatever ventilation and breathing pattern seems to fit the ventilatory control
system. PAV follows the equation of motion, which states that the pressure applied
by the respiratory muscles to the system, is used to overcome the elastic and
resistive opposing forces. The former is proportional to the volume displacement
whereas the latter is proportional to the airflow rate, neglecting inertia.
In theory, PAV should normalise the neuro-ventilatory coupling by making the
ventilator an extension of patient's respiratory muscles, while leaving to the
patient the entire control of all aspects of breathing. PAV, however, shares
a common problem with the conventional partial ventilatory support modes. In
mechanically ventilated patients, the respiratory system impedance may change
over time. These changes may impair the good matching between ventilator output
and patient's ventilatory demand and lead to patient-ventilator asynchrony.
The necessity of regular measurements of respiratory system mechanics and adjustments
in ventilatory setting, however, imposes a major obstacle to the widespread
use of this mode. (See Graph)
To overcome these problems, an ideal ventilator should be
able to record the activity of the respiratory neural system, and use that measurement
to select a satisfactory tidal volume, but at this time, it is not feasible
to record the activity of the respiratory centres in patients.
NAVA: Latest Module
The newly introduced NAVA, with the recording of the electrical activity of
the diaphragm, comes close. NAVA provides pressure assistance in proportion
to the electrical activity of the diaphragm, thereby ensuring a positive relationship
between the ventilator assistance and the patient's spontaneous effort. The
conventional ventilators have a pneumatic trigger and they deliver ventilatory
assist much after the diaphragmatic activity and this leads to increased effort
by the patient and ventilator-patient asynchrony whereas NAVA uses the electrical
activity of the diaphragm (EAdi) during spontaneous breathing to trigger, cycle
off and adjust the intra-breath assist profile in proportion to the EAdi. This
is done with the help of a special electrode, 12 cms long which is placed over
an esophageal catheter and inserted like a nasogastric tube. The signals coming
from these electrodes undergo a computer based analysis to adjust the ventilatory
assist for the delivered breaths. Signals coming from cardiac pulsation and
esophagus are filtered and then the remaining signals are amplified into a processing
unit. These signals are finally delivered to the ventilator, which finally assists
the patient proportional to the strength of the EAdi and a fixed gain constant.
A user-adjustable proportionality factor (the gain) is available to set the
amount of ventilator assistance for a given level of EAdi. EAdi represents the
global activity of the diaphragm and so the intactness of the phrenic nerve
is necessary for the reliability and strength of the signal. (see diagram).
The important advantages of NAVA are that it offers better
patient-ventilator synchrony, off-loading of inspiratory muscles, better response
time and the tight coupling between neural output and ventilator function allows
the ventilatory assist to be independent of changes in lung and ribcage elasticity,
flow resistance, intrinsic Positive-End-Expiratory Pressure (PEEP), leakage,
or abdominal compliance. Due to these properties, NAVA can also be used effectively
in delivering non-invasive ventilation as an alternative to endotracheal intubation
in hypoxemic respiratory failure.
Drdeepak_govil@yahoo.co.in
|